High temperature cooking system

The use of high-temperature steam to kill pathogenic microorganisms in medical waste is the main link of this medical waste high-temperature steam disposal center

SYSTEM CHARACTERISTICS
The sterilization room adopts a pulsating vacuum form for vacuum extraction, with a vacuum degree of 0.08MPa. The cycle of vacuum extraction and steam filling is not 4 times
The sterilization room adopts a pulsating vacuum form for vacuum extraction, with a vacuum degree of 0.08MPa. The cycle of vacuum extraction and steam filling is not 4 times
The final achieved indoor air discharge rate for sterilization is 98.5%. After completing the vacuum pumping, ensure that the vacuum degree of the sterilization chamber does not decrease by more than 1.3kPa (10mmHg) within 10 minutes
The final achieved indoor air discharge rate for sterilization is 98.5%. After completing the vacuum pumping, ensure that the vacuum degree of the sterilization chamber does not decrease by more than 1.3kPa (10mmHg) within 10 minutes
The B-D test package is used to test the vacuum pumping and steam penetration performance of the sterilizer before the first sterilization or after installation and maintenance every day, and it is an empty load test
The B-D test package is used to test the vacuum pumping and steam penetration performance of the sterilizer before the first sterilization or after installation and maintenance every day, and it is an empty load test